OpenAlex · Aktualisierung stündlich · Letzte Aktualisierung: 07.05.2026, 05:10

Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.

Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.

2006·397 Zitationen·PubMed
Volltext beim Verlag öffnen

397

Zitationen

9

Autoren

2006

Jahr

Abstract

UNLABELLED: PET combined with CT has proven to be a valuable multimodality imaging device revealing both functional and anatomic information. Although PET/CT has become completely integrated into routine clinical application and also has been used in small-animal imaging, CT provides only limited soft-tissue contrast and, in preclinical studies, exposes the animal to a relatively high radiation dose. Unlike CT, MRI provides good soft-tissue contrast even without application of contrast agents and, furthermore, does not require ionizing radiation. METHODS: This project focused on combining a high-resolution PET scanner with a 7-T MRI system for animal research. Because classic PET detectors based on photomultiplier tubes cannot be used in high magnetic fields, we used a detector technology based on 10 x 10 lutetium oxyorthosilicate crystal arrays and 3 x 3 avalanche photodiode arrays. A ring of such PET detectors will ultimately be used as an insert for the 119-mm-diameter MRI bore. RESULTS: Initial measurements with 1 PET detector module in the 7-T field during application of MRI sequences were encouraging. Position profiles from the PET detectors and a first MR image of a mouse could be acquired simultaneously. CONCLUSION: Further work will concentrate on the construction of a full PET detector ring with compact, integrated electronics.

Ähnliche Arbeiten

Autoren

Institutionen

Themen

Medical Imaging Techniques and ApplicationsRadiation Detection and Scintillator TechnologiesAdvanced X-ray and CT Imaging
Volltext beim Verlag öffnen